Safe and environmentally-friendly for sensitive instruments

Canon Lifecare Solutions newly developed low-temperature sterilization system with vaporized hydrogen peroxide technology. ES-1400 has effective 142L chamber capacity and it has 2 types of option, single or double-door, to meet your needs.

Features

Usable 142L Rectangular Chamber

Highly Productive 142 Liter chamber with rectangular shape can maximize productivity; it can fit full size containers and large size trays. It comes with 4 sliding racks which enables users to maximize the available chamber space in one cycle. Sliding racks help operators to remove the loads easily. The height can be adjusted depending on the size of medical instruments.

Full Colour LCD / LED Lighting

The full-colour touch panel LCD is designed for excellent ease of operation. The display would indicate errors with large texts on red background while alarming sound alerts users to take required immediate actions. LED lighting and window is convenient for increasing visibility inside the chamber. As the interior of the sterilization chamber is lighted on with white LED lamps, users can see how the items are laid out and if any items are missing though they intended.

Hands-free Operation

The door swings down automatically by either operation the touch panel or footswitch.

Consumables for ES-1400

Chemical Indicators

  • e-Card
Sterilization Bags
  • e-TR 70
  • e-TR 100
  • e-TR 150
  • e-TR 200
  • e-TR 250
  • e-TR 300
  • e-TR 350
  • e-TR 400
  • e-TR 450
  • e-TR 500
Sterilization Cartridge
  • e-SC102
  • e-SC3
  • e-SC305
  • e-SC610
Thermal Printer Paper
  • e-RP

    Specifications

    Specifications

    Product Name Hydrogen Peroxide Gas Sterilizer ES-1400
    Chamber Capacity 142L
    Process Time Normal Cycle
    Approx. 50 minutes
    Short Cycle Approx. 28 minutes
    Door Type Automatic Swing-Down Door
    Double or Single-Door (Option)
    Sterilization Mode Standard, Soft
    Sterilization Temperature 55°C
    Display Full Colour LCD Touch Panel
    Printing / Recording Line Thermal Printer / SD Card
    Power Requirement 3-phase 208V / 380V / 400V / 415V, 50 / 60 Hz, 3kW
    Required Space (mm)* Left: +50, Right: + 150, Back: +50, Top: +500
    Operational Environment Temperature:? 10 - 40°C
    Humidity: 30 - 85% RH
    External Dimensions
    (Wx D x H)
    750 x 925 x 1710mm
    Weight Approx. 410 kg (Single-Door Type)

    Product Disclaimer